Motion: Automated, stepper-driven X-axis travel of at least 9 meters and Z-axis travel extending at least 30 cm below beam centerline; 1 mm positioning accuracy and repeatability of 1 mm or better on both axes.
Instrument platform: 50 × 50 cm with a 1 cm engraved grid, tapped mounting holes at 10 cm intervals, and at least 100 lb capacity with no measurable deflection.
Irradiator stand: Fixed stand with at least 250 lb capacity and beam height of 1.0–1.5 meters above the floor; mounting and alignment compatible with existing IDNS cradle and JLS irradiator drawings.
Controls: Console or PC interface with commanded and actual positions, 1 mm display resolution, X-axis instrument offset, smooth travel without overshoot, limit switches, homing, and safety interlocks.
Construction: Rigid structure suitable for laboratory use near radiation fields, full-travel cable management, and M6x1.0 tapped holes.
Delivery and installation: Include shipping, onsite installation, acceptance testing, labor, equipment, and travel. Components must fit through a nominal 48-inch doorway; installation must be complete by June 30, 2027.
Documentation and warranty: Mechanical drawings, dimensions, operation and maintenance manual, spare parts list, and transferable standard manufacturer's warranty.